distilation & filtration & separation & sample preparation
171
Offers ideal solutions for determination of erithrocytes,
determination of the biluribin content and extraction of DNA
and RNA.
Explosion proof and marked as CE, cTUVus, and FCC. MCA
test according to IEC/EN61010-2-10.
Passed EN610-2-101:2002 particular requirements for in
vitro diagnostic (IVD) medical equipment.
The unique design guarantees stable running and high
accuracy of the speed. High strength rotor and excellent
balance technology guarantees a quiet and stable operation.
DC motor accelerates the rotor to the set speed with low
vibration and a quiet operation less than 60 dB noise level.
User friendly large LCD display enables easy setting and
reading of the funtional parameters.
Starts and reaches a max speed of 14.000 rpm with max
18.620 RCF in seconds. Speed can be set with an accuracy
of +/- 30 rpm.
Metal casing with low gravity, provides stable and reliable
running at maximum speed.
Timer can be set from 30 seconds to 99 minutes or for
continuous running. Timer starts once the target speed is
reached for more accurate operation time.
Short spin function is used for quick spins by pressing and
holding the “pulse” key.
Multi voltage design guarantees stable running and high
accuracy of speed.
Featured with a dual door interlock system for twice safety.
Electronic brake immediately stops when lid opens for user
safety. Lid is released automatically when the rotor stops.
CENTRIFUGE -
“hematocrit”
catalogueno
603.06.001
maximumspeed (rpm)
voltage (V)
noise level (dB)
dimensions /weight (mm/ kg)
maximumRCF (xg)
speedaccuracy (rpm)
increment (rpm)
frequency (Hz)
run time
acceleration / braking time (sec)
additional features
safetydevices
280x364x266 / 12
220
50 / 60
14000
18620
± 30
10
< 60
30 sec. - 99 min. or continuous
20 s 20 s
door interlock
over - speed detection
over - temperature detection
automatic internal diagnosis
automatic rotor identification
sound alert function
speed / RCF swich
short time run function
24 cavity rotor for capillary tubes.
catalogue no: 603.06.101
24 cavity rotor for 2 ml/ 1,5 ml and
0,5 ml tubes and related tube inserts.
catalogue no: 603.06.102
User friendly large LCD display offers
easy identification of parameters
and shows all informations including
accumulated running time and last
running parameters.
Both RPM or G-force can be set
and displayed. Parameters can
be modified after the set speed is
achieved.
Quick spin function is used for quick
spins by pressing and holding the
“pulse” key.
Supplied without rotor. Rotor should be ordered separately
High strength plastic rotor and excellent balance technology to
guarantee a quiet and stable operation.