Previous Page  191 / 482 Next Page
Information
Show Menu
Previous Page 191 / 482 Next Page
Page Background

distilation & filtration & separation & sample preparation

171

Offers ideal solutions for determination of erithrocytes,

determination of the biluribin content and extraction of DNA

and RNA.

Explosion proof and marked as CE, cTUVus, and FCC. MCA

test according to IEC/EN61010-2-10.

Passed EN610-2-101:2002 particular requirements for in

vitro diagnostic (IVD) medical equipment.

The unique design guarantees stable running and high

accuracy of the speed. High strength rotor and excellent

balance technology guarantees a quiet and stable operation.

DC motor accelerates the rotor to the set speed with low

vibration and a quiet operation less than 60 dB noise level.

User friendly large LCD display enables easy setting and

reading of the funtional parameters.

Starts and reaches a max speed of 14.000 rpm with max

18.620 RCF in seconds. Speed can be set with an accuracy

of +/- 30 rpm.

Metal casing with low gravity, provides stable and reliable

running at maximum speed.

Timer can be set from 30 seconds to 99 minutes or for

continuous running. Timer starts once the target speed is

reached for more accurate operation time.

Short spin function is used for quick spins by pressing and

holding the “pulse” key.

Multi voltage design guarantees stable running and high

accuracy of speed.

Featured with a dual door interlock system for twice safety.

Electronic brake immediately stops when lid opens for user

safety. Lid is released automatically when the rotor stops.

CENTRIFUGE -

“hematocrit”

catalogueno

603.06.001

maximumspeed (rpm)

voltage (V)

noise level (dB)

dimensions /weight (mm/ kg)

maximumRCF (xg)

speedaccuracy (rpm)

increment (rpm)

frequency (Hz)

run time

acceleration / braking time (sec)

additional features

safetydevices

280x364x266 / 12

220

50 / 60

14000

18620

± 30

10

< 60

30 sec. - 99 min. or continuous

20 s 20 s

door interlock

over - speed detection

over - temperature detection

automatic internal diagnosis

automatic rotor identification

sound alert function

speed / RCF swich

short time run function

24 cavity rotor for capillary tubes.

catalogue no: 603.06.101

24 cavity rotor for 2 ml/ 1,5 ml and

0,5 ml tubes and related tube inserts.

catalogue no: 603.06.102

User friendly large LCD display offers

easy identification of parameters

and shows all informations including

accumulated running time and last

running parameters.

Both RPM or G-force can be set

and displayed. Parameters can

be modified after the set speed is

achieved.

Quick spin function is used for quick

spins by pressing and holding the

“pulse” key.

Supplied without rotor. Rotor should be ordered separately

High strength plastic rotor and excellent balance technology to

guarantee a quiet and stable operation.